Mode and Programme
Selection
1. Press the mode selection button to
scroll through and select the desired
mode.
2. Once you have selected the mode,
press
the
programme
button to scroll through and select the
programme of choice.
Each time you change the mode or
programme, the intensity level will revert
to zero. This is a safety feature to alleviate
any sudden feeling of a surge, as each
programme gives a different sensation.
Selecting Treatment Time
The default treatime time for preset
programmes is shown. For relevant
programmes, treatment time can be
changed by pressing the settings button.
Use CH2 ▲ ▼ to select the preferred
treatment time. You can select between
5 - 90 minutes or continuous (marked as
C).
Setting the Intensity
Once the electrode pads have been
connected and applied to the body and
the programme has been selected, the
intensity levels can be adjusted. Unipro
has 99 levels of intensity.
CH1 ▲ CH1 ▼ control the intensity of the
left-hand lead/electrode pads. CH2 ▲
CH2 ▼ control the intensity of the right-
hand lead/electrode pads. Each press
of the ▲ and ▼ buttons will adjust the
intensity by 1mA. The intensity can be
10
increased
pressing and holding the ▲ ▼ buttons.
At low intensity levels, it is common for no
sensation to be felt. Continue to increase
the intensity until the sensation is strong
but comfortable.
During a session, your body can beome
used to the sensation. If this is the case,
selection
increase the intensity levels.
The intensity controls will not operate
unless the Unipro has been properly
conencted to the body. If the Unipro
detects
channel, the intensity will rest to 0mA and
the connection error icon will flash. This
is a safety feature designed to prevent
sudden surges in stimulation.
When a green circle is displayed next to
the intensity levels, this means the output
is high enough be harmful if not used
correctly. Please use with caution.
Manual Setting Selection
In most cases, the programmes settings
are
identified by 'P'. When a programme has
manual settings available, 'M' will be
displayed.
In these programmes you are able
to select and change the following
parameters:
Frequency (Hz)
Pulse Width (μs)
Treatment Time (min)
Work Time (EMS programmes only) (s)
Rest Time (EMS programmes only) (s)
or
decreased
a
disconnection
preset.
These
programmes
quickly
by
on
either
are