Control
Closing
Opening
Gripping force
6.2.5 Programme 5: Digital Control
Proportional control with 2 electrodes or a switch.
The terminal device opens and closes at the fastest speed. The gripping force increases propor
tionally to the duration of gripping.
Control
Closing
Opening
Gripping force
Speed
6.2.6 Programme 6: Double Channel Control
Control with 1 electrode or 1 linear control element.
The terminal device closes with a slow, weak signal and opens with a fast, strong signal. When
one electrode is used, the upper switching threshold can be adjusted with the 757T13 MyoSe
lect.
Control
Closing
Opening
Gripping force
Speed
2 electrodes, proportional
Proportional to the decrease in the strength and speed of the sig
nal on electrode 1.
Proportional to the increase in the strength and speed of the signal
on electrode 1.
Proportional to the signal strength on electrode 2.
2 electrodes or switch, digital
Constant speed with detection of the "OPEN" signal on electrode
1 or the switch
Constant speed with detection of the "CLOSE" signal on elec
trode 2 or the switch
Proportional to the signal duration on the "CLOSE" electrode or
switch. The longer the signal is applied, the higher the gripping
force.
Constant speed of 200 mm/s for opening and closing
1 electrode
Slow, weak electrode signal that does not reach the upper
threshold within 80 ms.
Quick, strong electrode signal that reaches the upper threshold
within 80 ms.
Proportional to the signal duration. The longer the signal is
applied, the higher the gripping force
Constant speed of 200 mm/s for opening and closing
21