Operation
7.3.2 Polytron 8100 EC-specific symbols
Symbol
Explanation
The sensor is ready for operation.
The sensor is ready for operation but is
close to the end of its useful life.
The sensor is still ready for operation but
should be changed as soon as possible.
The vitality is <25%.
LT
A LifeTest of the Dräger SensorAlive is car-
ried out.
7.3.3 Measuring mode
The following display examples show the Polytron 8100 EC.
Gas names are not predefined for the Polytron 8200 CAT and
Polytron 8310 IR. Gas names must be entered via the
DrägerPolysoft PC software. After modifying parameters with
the DrägerPolysoft PC software or another software, verify all
parameters by downloading them from the Polytron® 8xx0 or
by checking them directly on the instrument.
The following is displayed in measuring mode:
Display example
Description
In normal operation, the display
shows the measured gas concentra-
20.9
tion, the selected gas type, and the
unit of measurement.
Analog interface: Corresponding
O 2
Vol.%
current value
Relays: All relays are in the normal
state
Exceeding the alarm thresholds.
Example shows pre-alarm A1.
19.0
Analog interface: Corresponding
current value
Relays: Relay A1 is in alarm state.
O 2
A1
7.3.4 Special states
If the instrument is in a special state, proper measurement or
alarm activation can no longer be guaranteed.
This is the case:
– When the gas concentration exceeds the measuring
range.
– When a fault occurs.
– When alarms are deactivated.
– When calibrations are performed.
– When bump tests are performed.
– When a sensor is in the warm-up phase.
– When the system is in maintenance mode.
– When a Dräger SensorAlive is used and the system is
configured such that the measured value is not indicated
during a LifeTest.
For the current of the analog interface corresponding to the
analog signal, see 16.2.
16
The signals of the digital interfaces are described in the
following documents.
Interface
PROFIBUS PA
Foundation Fieldbus FF
Modbus RTU
Display example
Description
Value above measuring range
The gas concentration is outside of
the measuring range of the sensor
Analog interface: Over range
O 2
A2
Relays: A2 relay switch
Note for the Polytron 8200 and
Polytron 8310:
The message that the measuring
range has been exceeded can also be
caused by a short circuit or an inter-
ruption in the line to a remote sensor.
Value below measuring range
The gas concentration is outside of
the measuring range of the sensor.
Analog interface: Drift below zero
Relays: Fault relay switch
O 2
Vol.%
Fault display
The
side of the display.
Analog interface: Fault current
Relays: Fault relay switch
Maintenance display
20.9
The
side of the display.
This is displayed when alarms are
deactivated, and during adjustment,
O 2
Vol.%
bump tests, and maintenance work.
Analog interface: Maintenance cur-
rent
Relays: No change
Warning display
20.9
The
side of the display.
Analog interface: Warning current
O 2
Vol.%
Relays: No change
Warm-up phase 1
The
Sensor ready
side of the display.
in 0:09:13
The remaining time is displayed on
O 2
Vol.%
the left side.
Analog interface: Maintenance cur-
rent
Relays: Fault relay switch
The fault relay behavior can be con-
figured for warm-up phase 1 (see
"Setting fault relay warm-up 1",
page 30).
Instructions for use
Document
9033782
9033783
9033781
symbol is displayed on the right
symbol is displayed on the right
symbol is displayed on the right
symbol is displayed on the right
®
|
Dräger Polytron
8000 Series
1)
.
2)