13
Sensor settings
13.1 Switching automatic calibration on or
off
1. Select Settings > Sensor > Auto calibration and
confirm.
2. Select Enable or Disable confirm with [OK].
13.2 Capture range
The capture range blanks out measurement fluctuations.
Measurement fluctuations are minor variations in measured
values (such as signal noise, variations in concentration).
Those variations do not change the transmitted or displayed
value. Measured values within this range will be displayed
with the capture value. Measured values outside the capture
range will be displayed with the actual measured value.
Capture value (offset)
The capture value is displayed continuously, as long as the
measured value ranges between the upper and lower capture
value limits.
Upper capture value limit
The upper capture value limit marks the upper limit of the
range in which the capture value is displayed.
Lower capture value limit
The lower capture value limit marks the lower limit of the
range in which the capture value is displayed.
13.2.1 Setting capture range
1. Select Settings > Sensor > Display Capture and
confirm.
2. Set offset in the first window.
3. Set the lowest value of the capture range and confirm with
Next.
4. Set the highest value of the capture range and confirm
with Next.
5. Check all the values and confirm with Next.
13.2.2 Switching capture range off
1. Select Settings > Sensor > Display Capture and
confirm.
2. Set offset in the first window.
3. Set the lowest value of the capture range to 0 and confirm
with Next.
4. Set the highest value of the capture range to 0 and
confirm with Next.
5. Check all the values and confirm with Next.
13.3 Resetting sensor
This function resets the sensor to factory default settings.
1. Select Settings > Sensor > Init. channel and confirm.
2. Select Confirm and confirm with [OK].
Instructions for use
|
Dräger Polytron
®
8000 Series
13.4 Setting calibration interval
After the calibration interval expires, the instrument will issue
a warning that the calibration is past due.
1. Select Settings > Sensor > Cal. interval and confirm.
2. Set the calibration interval and confirm with [OK].
13.5 Setting sensor lock
This point is only valid for Polytron 8100 EC and Polytron
87x0 IR.
1. Select Settings >Sensor > Sensor lock and confirm.
2. Select On or Off.
On
The instrument rejects any sensor whose
part number does not match the part num-
ber of the previously installed sensor.
The instrument accepts any suitable sensor,
Off
uploading the default settings of this sensor.
However, this overwrites any customized
settings with the default settings of the new
sensor.
13.6 Software dongles for EC sensors
13.6.1 Installing the dongle
1. Switch off power to the instrument or declassify the area
according to the local regulations.
2. Loosen set-screw and unscrew lid from instrument.
3. Insert the dongle with the Dräger logo facing upwards in
the slot of the PCB unit, See figure D on the fold-out
page..
4. Position the PCB unit back cover and close the instrument
(see "Closing the gas detector", page 13).
13.6.2 Deactivating dongles
This function deactivates a dongle to safely remove the
dongle or if the dongle is faulty.
A dongle can only be reactivated by cycling the power to the
instrument.
1. Select Settings > Instrument > SW dongle > and the
dongle to be deactivated, and confirm.
2. Select Deactivation func.. The selected dongle is
deactivated.
13.7 Sensor settings Polytron
13.7.1 Sensor self-test
This function is only active if the sensor-test dongle or
diagnostic dongle is installed. If the sensor does not pass the
self-test, a warning or error message is issued.
Setting regular sensor self-tests
This feature triggers the sensor self-test at regular intervals.
The instrument then routinely checks that the sensor is
functioning correctly.
1. Select and confirm Settings > Sensor > Set Sensortest.
2. Select and confirm Enable or Disable.