Disposal
15
Disposal
This product must not be disposed of as household
waste. This is indicated by the adjacent icon.
You can return this product to Dräger free of charge. For
information please contact the national marketing
organisations and Dräger.
Disposal of electrochemical sensors
WARNING
Danger of explosions and risk of chemical burns!
Sensor fluids may leak out and can cause acid burns.
► Do not dispose sensors in fire.
► Do not open with force.
16
Technical data
16.1 Measuring ranges
Sensor
Polytron 8100 EC with
®
DrägerSensor
EC
Polytron 8200 CAT with
®
DrägerSensor
DQ
Polytron 8200 CAT with
®
DrägerSensor
LC
Polytron 8310 IR with Dräger-
®
Sensor
IR
Polytron 8700 IR with
PIR 7000 infrared gas sensor
Type 334 (IDS 01x1)
Type 340 (IDS 01x2)
Polytron 8720 IR with
PIR 7200 infrared gas sensor
(IDS 01x5)
1) For adjustable full scale deflection, see 14.3.
2) Range changes depending on the substance to be measured
16.2 Signal transmission to the controller
The measuring range and performance characteristics
depend on the installed sensor (see the instructions for use
for the installed sensor).
Frequency of measurement calculation: 1 x per second
(update of display, analog interface and relays).
42
Measuring range
Depending on sensor
0 to 100 %LEL
1)
0 to 10 %LEL
0 to 100 %LEL
1)
0 to 10,000 ppm
1)
0 to 100 %LEL,
0 to
1)2)
(1.25 x LEL
) Vol%,
max
For CH4 (methane) 0 to
1)
100 Vol%
1)
0 to 10,000 ppm
1)
0 to 100 %LEL
1)
0 to 5 Vol%
1)
0 to 100 Vol%
Analog signal
Normal operation
Drift below zero
Over range
Instrument fault
Fault on analog
interface
Maintenance signal
Resistance of the signal loop
Operating
Resistance
mode
range of the
signal loop
Operation
0 to 230 Ω
without
Rising linearly
®
HART
com-
with the supply
munication
voltage from:
0 to 230 Ω at
10 V to
0 to 500 Ω at
16 V
0 to 500 Ω
Operation
230 to 270 Ω
®
with HART
Rising linearly
communica-
with the supply
tion
voltage from:
230 to 270 Ω at
11 V to
230 to 500 Ω at
16 V
230 to 500 Ω
230 to 500 Ω
®
HART
mul-
tidrop opera-
tion
When used in accordance with the metrological suitability test
compliant with EN IEC 62990-1, EN IEC 60079-29-1 or
EN 50104, the resistance must not exceed 300 ohms. Further
information on installing the fieldbus interface is provided in
document 9033783.
Instructions for use
4 ... 20 mA
3.8 ... 4 mA
20 ... 20.5 mA
≤ 1.2 mA
> 21 mA
3.4 mA steady signal or 1 Hz mod-
ulation between 3 and 5 mA
(selectable)
Power supply range
at 10 V DC
10 to 18 V DC
18 to 30 V DC
at 13 V DC
11 to 16 V DC
18 to 30 V DC
10 to 30 V DC
®
|