4.6 Protective function of automatic reverse
During operation, if the load value exceeds the preset torque value, the fi le
rotation mode will automatically change to Reverse Mode. And the fi le would
return to normal rotation mode when the load is below the preset torque value
again.
Cautions:
1. Protective function of automatic reverse is ONLY suitable for CW mode.
Press Setting button "P" again, the "Dominant Hand" can be
changed, press Adjusting button "+"/"-" to adjust, then press to
"Main" button to confi rm. The right hand and the left hand can
be set.
Press Setting button "P"again, the "Calibration" can be
changed, press Adjusting button "+"/"-" to select "ON", then
press "Main" button for calibration. Before calibrating, make
sure the original contra angle is installed, and do not install the
fi le.
The torque will not be correct if it is calibrated without original
contra angle or if there is any load on contra angle chuck, and
there is risk of fi le damage.
After replacement of contra angle, the contra angle should be
calibrated before use.
Press Setting button "P"again, the "Beeper Volume" can be
change, press Adjusting button "+"/"-" to adjust, then press to
"Main" button to confi rm.
The"Beeper Volume"can be set from 0-3. Vol.0: Mute.
Press Setting button "P" again, the "Restore Defaults" can be
change, press Adjusting button "+"/"-" to select "ON", then
press to "Main" button to restore defaults.