use the original lithium battery and replace the lithium battery according to
the correct steps in the instructions.
1.7.14 Please remove the battery if the motor handpiece is not likely to be
used for some time.
1.7.15 Wireless charging will generate heat, and the surface temperature
of charging base and motor handpiece will rise. It is recommended not to
come in contact with the motor handpiece and charging base during wireless
charging for more than 10 seconds.
1.8 Device safety classifi cation
1.8.1 Type of operation mode: Continuous operating device
1.8.2 Type of protection against electric shock: Class II equipment with
internal power supply
1.8.3 Degree of protection against electric shock: B type applied part
1.8.4 Degree of protection against harmful ingress of water: Ordinary
equipment (IPX0)
1.8.5 Degree of safety application in the presence of a fl ammable anesthetic
mixture with air, oxygen, or nitrous oxide: Equipment cannot be used in the
presence of a fl ammable anesthetic mixture with air, oxygen, or nitrous oxide.
1.8.6 The contact duration of applied part: 1 to 10 minutes.
1.8.7 Applied part: contra angle, lip hook,fi le clip,touch probe.
1.8.8 The temperature of the surface of applied part may reach 46.6℃.
1.9 Primary technical specifi cations
1.9.1 Battery
Lithium battery in motor handpiece: 3.7V /2000mAh
1.9.2 Power adapter (Model: ADS-6AM-06N 05050/ UE08WCP-050100SPA)
Input: ~100V-240V 50Hz/60Hz, 0.4A
Max Output: 5V
1.9.3 Torque range: 0.4Ncm-4.0Ncm(4mNm ~ 40mNm)
1.9.4 Speed range: 100rpm~3000rpm
1.9.5 Wireless charging Frequency range: 112-205KHz
Maximum RF output power of the product: 9.46dBuA/m@3 m
1.10 Environment parameters
1.10.1 Environment temperature: +5℃ ~ +40℃
1.10.2 Relative humidity: 30% ~ 75%
1.10.3 Atmospheric pressure: 70kPa ~ 106kPa