Technical Description Concerning Electromagnetic Emission
Table 1: Declaration - electromagnetic emissions
Guidance and manufacturer's declaration - electromagnetic emissions
The model EndoPro Ai1 is intended for use in the electromagnetic environment specifi ed
below.
The customer or the user of the model EndoPro Ai1 should assure that it is used in such
an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR11
Harmonic emissions
lEC 61000-3-2
Voltage fl uctuations
/ fl icker emissions
lEC 61000-3-3
Technical Description Concerning Electromagnetic Immunity
Table 2: Guidance & Declaration - electromagnetic immunity
Guidance & Declaration — electromagnetic immunity
The model EndoPro Ai1 is intended for use in the electromagnetic environment specifi ed
below.
The customer or the user of the model EndoPro Ai1 should assure that it is used in such
an environment.
Immunity test
Electrostatic
discharge
(ESD)
lEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
lEC 61000-4-5
Compliance
The model EndoPro Ai1 energy only for its internal
function. Therefore, its RF emissions are very
Group 1
low and are not likely to cause any interference in
nearby electronic equipment.
Class B
The model EndoPro Ai1 is suitable for used in all
establishments, including domestic establishments
Class A
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Complies
IEC 60601
test level
±8kV contact
±2, ±4, ±8, ±15kV
air
±2kV for power
supply lines
±1kV for Input/
output lines
±0.5, ±1kV line to
line
±0.5, ±1, ±2kV
line to earth
Electromagnetic environment - guidance
Compliance level
±8kV contact
±2, ±4, ±8, ±15kV
air
±2kV for power
supply lines
±0.5, ±1kV line to
line
±0.5, ±1, ±2kV
line to earth
35
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If fl oors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.