It is recommended to contact local distributors or manufacturer to replace the
battery.
4.10 Oiling of contra angle
Only the original oil injection nozzle can be used for oiling of contra angle.
The contra angle needs to be lubricated after cleaning and disinfection, but
before sterilization.
1. Firstly, screw the injecting nozzle into jet of oil bottle. (Around 1 to 3
circles).
2. Next, plug the nozzle into the end part of contra angle, and then grease the
contra angle for 2-3s till the oil fl ow out of contra angle head part.
3. Place the contra-angle upright for 30 minutes for leftover oil to fl ow out.
Warnings
Motor handpiece cannot be fi lled with oil.
Cautions
a: To avoid the contra angle from fl ying due to pressure, use hand to safely
hold the contra angle while lubricating.
b: Please use the appropriate nozzle suitable for handpiece lubrication.
5 Troubleshooting
Failure
The motor handpiece does
not rotate.
There is continuous beep
sounds after starting the
motor handpiece.
Contra angle calibration
failure
Possible cause
Motor is on EAL mode,
EAL mode is only for canal
measurement.
The continuous beep sound
is indicating that the motor
handpiece is under CCW mode.
Calibration failure caused by
strong resistance of contra angle
24
Solutions
Change to CW, CCW, SGP,
CW-ATR or CCW-ATR.
Stop the motor handpiece and
change the operating mode to
CW Mode.
Clean the contra angle, and
recalibrate after oil injection.