EP AI1 TOUCH PROBE
EP AI1 HEAD COVER 3PCS
EP AI1 POWER ADAPTER
2.2 Display Screens
2.2.1 Display Screens for 6 Operation Modes and Standby
2.2.1.1 EAL Mode
This mode is for canal measurement. The motor handpiece does not run in
this mode.
2.2.1.2
CW Mode
The motor handpiece rotates forward 360º, clockwise direction.
2.2.1.3 CCW Mode
The motor handpiece rotates counterclockwise direction only. This mode
is used to inject calcium hydroxide and other medicant. When this mode is
being used, a double-beep sounds continuously.
2.2.1.4 SGP Mode Safety Glide Path Mode
F: Forward angle, R: Reverse angle
The rotation angle is adjustable, but the forward angle must be equal to the
reverse angle.
2.2.1.5 CW-ATR Mode and Adaptive Torque Reverse function.
Normal continuous forward rotation, when the load of the file is greater than