Make sure of proper contact between socket and measuring wire
Remove the dentin chippings or residual pulp
Keep pushing endo file near to apical foramen, screen display will become normal
Reduce flush fluid, avoid contacting metal prosthesis
Completely stop bleeding and clean the blood
Insulate the endo file and dental crown with insulator or cut off the gum hyperplasia
Clean necrotic tissue of the proximal caries, if it's the same with adjacent dental
surface and periodontium, temporarily the adjacent surface
With help of X ray film, avoid touching ledge with endo file
Remove the gutta-percha or debris and moist the root canal
Solutions
Solutions
Replace the file clip
Replace the measuring wire
Use the endo file with larger diameter
Rinse root canal with saline solution
Remove the gum
Moist the root canal with saline solution.
Avoid endo file touching metal prosthesis
Reconnect the lip hook or replace lip hook
Avoid endo file touching metal prosthesis
Combine X-ray film to get the length
Remove residue, metal residue and debris
Repair perforation
Replace file clip
Rinse root canal with saline solution
Moist the root canal with saline solution