3. Instructions for aspiration operation on the patient:
3.1.Connect the accessories, referring to the 'CONNECTION DIAGRAM' on the cover.
ATTENTION: the filter is hydrophobic and in addition to being antibacterial/antiviral, it also acts as a secondary protection and
stops fluids that the primary protection device in the lid of the collection vessel accidentally fails to stop.
Follow the instructions below for the correct assembly of the filter:
Lato da collegare alla presa - Side to be connect-
VACUUM (12) del vaso di raccolta - VACUUM of
the collection vessel
3.2. Operate the appliance according to the indications of the selected operation. Always use the appliance on a horizontal plane;
this will allow the liquid ingress protection device in the aspiration pump to function properly.
3.3. For more comfortable aspiration, set the desired vacuum value (bar) using the vacuum regulator (4). Turning the knob clock-
wise gives a higher vacuum value and turning it counterclockwise gives a lower vacuum value; these values can be read on the
"vacuum gauge" (3).
3.4. Place your finger on the hole of the manual aspirated flow control (15) and, in pulsed mode, start aspiration on the patient
through the cannula (NOT INCLUDED WITH THE DEVICE).
3.5. When the application is finished, switch off the appliance.
3.6. Disconnect the connecting pipes and if necessary close the 'PATIENT' and 'VACUUM' outlets with plugs 14a -14b. During aspi-
ration, place caps 14a and 14b on fixing anchors 14d and 14c.
3.7. Empty and clean the vessel and connecting pipes.
14b
14d
ed to the socket
14c
14a
Writing:
Lato aspiratore - Aspirator side
Lato da collegare alla presa - Side to be
connected to the socket
ENTRATA ARIA (5) dell'aspiratore - AIR INLET
of the aspirator
14a
14b
23

Publicidad

loading

Este manual también es adecuado para:

Aspira go p1611em-30