(method A): 134°C and a time of 10 minutes first.
(method B): 121°C and a time of 20 minutes first.
Storage:
After sanitising, disinfection or sterilisation, reassemble the vessel and connecting pipes according to the instructions given in the
'CONNECTION DIAGRAM' .
At the end of each use, store the medical device complete with accessories in a dry and dust-free place.
• The manual aspirated flow control (15) is a sterile disposable product and must be replaced each time it is used.
• The ANTIBACTERIAL HYDROPHOBIC FILTER (7) is a single-use product and must be replaced each time it is used.
ACCESSORY COMPARTMENT
The accessory compartment (17) can be easily removed for more thorough cleaning. See the disassembly sequence on page 1 in
drawings 8-9-10-11.
VESSEL HOLDER
For ease of use, the vessel holder can be disassembled and reassembled on the side (RIGHT or LEFT) considered most suitable. With
the use of a tool (draw.12) disengage the vessel holder from the bottom of the appliance (draw.13) and rotate it so that it disengages
(draw.14) Re-engage it, proceeding with the reverse operation, on the chosen side. Make sure that the vessel holder is properly hooked
to the bottom of the appliance. This system allows up to two vessel holders to be connected. (draw.15)
6
HYGIENIC PREPARATION IN THE HOME
method A
method B
method C
MAX 300 TIMES
HYGIENIC PREPARATION IN A CLINIC OR HOSPITAL
method A
MAX 50 TIMES
method B
\
✓: planned \: not planned
Table of planned methods / patient accessories
8
9
Sanitisation
Disinfection
MAX 300 TIMES
MAX 300 TIMES
Sterilisation
\
MAX 50 TIMES
\
MAX 50 TIMES
25
10a
10b
MAX 300 TIMES
MAX 300 TIMES
\
\
MAX 50 TIMES
MAX 50 TIMES
11
MAX 300 TIMES
\
MAX 50 TIMES