English
4.1
DC VOLTAGE MEASUREMENT
1.
Switch the calibrator on using the On/Off button (5).
2.
Press the input/output selection key (8) to display
"INPUT" (13).
3.
Press the V mV selection key (6) to select V DC or
mV
DC
measurement range.
4.
Connect the red test lead to the V terminal (4) and
the black lead to the COM terminal (3).
5.
Connect the other end of the red test lead to the
positive pole of the voltage to be measured and
the black lead to the negative pole (earth).
6.
The result is displayed on (16).
* The numbers in parenthesis refer to the Description of the front panel (Page 21)
and the Description of the display screen (Page 22).
4.2
DC VOLTAGE OUTPUT
1.
Switch the calibrator on using the On/Off button (5).
2.
Press the input/output selection key (8) to display
(14) "OUTPUT".
3.
Press the V mV selection key (6) to select V or mV
(20) according to the desired output voltage.
4.
Press the adjustment keys (9), (10), (11), (12), to
set the desired value.
5.
Connect the red test lead to the V terminal (4),
and the black lead to the COM terminal (3).
6.
Connect the other end of the red test lead to the
positive pole of the voltage to be measured and
the black lead to the negative pole (earth).
7.
If you want to change the output value or range,
press the setting adjustment keys (9), (10), (11),
(12) or the V mV selection key (6).
4 DIRECTIONS FOR USE
(20)
according
Voltage / current calibrator C.A 1631
to
the
desired
23
-
+