Descargar Imprimir esta página

Chauvin Arnoux 560CA1631 Manual De Instrucciones página 25

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

English
4.5
DC CURRENT OUTPUT
Source of current (generation).
1.
Switch the calibrator on using the On/Off button (5).
2.
Press the input/output selection key (8) to display
(14) "OUTPUT".
3.
Press the mA–mA% (7) selection key to select mA
or mA% (19) according to the desired display.
In the mA% mode, the 4-20 mA value will be
displayed in the secondary display area (22).
Press the value setting keys (9), (10), (11), (12), to
set the desired value.
4.
Connect the red test lead to the LOOP terminal (1),
and the black lead to the V terminal (4).
5.
Connect the other end of the red test lead to the
positive pole of the output current and the black lead
to the negative pole.
6.
If you want to change the output value, press the
value setting keys (9), (10), (11), (12) or the
selection key mA–mA% (7) to change the display.
4.6
SIMULATION OF TRANSMITTER
1.
Switch the calibrator on using the On/Off button (5).
2.
Press the input/output selection key (8) to display
(14) "OUTPUT".
3.
Press the mA–mA% (7) selection key to select mA
or mA% (19) according to the desired display.
In the mA% mode, the 4-20 mA value will be
displayed in the secondary display area (22).
4.
Press the value setting keys (9), (10), (11), (12), to
set the desired value.
5.
Connect the red test lead to the V terminal (4), and
the black lead to the COM terminal (3).
6.
Connect the other end of the red test lead to the
positive pole of the external power supply and the
black lead to the positive pole of the current to be
measured.
7.
If you want to change the output value, press the
setting keys (9), (10), (11), (12) or the mA–mA% (7)
selection key to change the display.
Voltage / current calibrator C.A 1631
25
-
+
A
+
-
+
MAX
power
supply
30
A
-

Publicidad

loading

Este manual también es adecuado para:

Ca1631