Voltage / current calibrator C.A 1631
7.2
AC/DC CHARACTERISTICS OF THE MAINS ADAPTER
Input: 100 V – 240 VAC, 50 – 60 Hz 1.8 A
Output: 12 VDC
Polarity:
WARNING:
1.
Use the original mains adapter. Other models might damage your instrument;
2.
The adapter is intended for indoor use only.
3.
First connect the plug of the AC cord to the mains outlet, then firmly insert the
supply plug into the DC jack of the measuring device. To disconnect it, pull the
DC plug straight out, then disconnect the adapter from the mains outlet.
4.
Do not use the mains adapter with any other device.
5.
It is normal for the mains adapter to warm up in operation.
6.
Do not dismantle the mains adapter. This would be dangerous.
7.
Do not use the adapter in an overheated or damp room.
8.
Avoid jolting the mains adapter.
9.
It is normal for the mains adapter to make noise when in operation.
For maintenance, use only the spare parts specified. The manufacturer
cannot be held liable for any accident that occurs following a repair not performed
by its customer service department or by an approved repairer.
8.1
CLEANING
•
Disconnect the unit completely and turn the rotary switch to OFF.
•
Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry
rapidly with a dry cloth or forced air. Do not use alcohol, solvents, or
hydrocarbons.
•
Make sure that no foreign body interferes with the operation of the snap device
of the sensor.
8.2
CALIBRATION
•
Calibrate your calibrator once a year to make sure that it functions in conformity
with the specifications.
2 A MAX
8 MAINTENANCE
30
English