Descargar Imprimir esta página

Chauvin Arnoux 560CA1631 Manual De Instrucciones página 31

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

English
8.3
REPLACING THE BATTERIES
Change the batteries when the LCD screen indicates
To change the batteries, disconnect all leads, switch the calibrator off, unscrew
the battery compartment cover, and replace all of the batteries with new 1.5V AAA
batteries.
8.4
REPLACING A FUSE
To avoid injuring a person or damaging the calibrator, use only fast-
blow 0.125 A, 250 V fuses.
Fuse 1 has probably blown if:
In the voltage output mode, with the test leads disconnected from the
calibrator, OL flashes on the screen.
Fuse 2 has probably blown if:
In the current input mode, the calibrator always displays 0.000, even when
a signal is applied.
8.5
METROLOGICAL CHECK
Like all measuring or testing devices, the instrument must be
checked regularly.
This instrument should be checked at least once a year. For checks and
calibrations, contact one of our accredited metrology laboratories (information and
contact details available on request), at our Chauvin-Arnoux subsidiary or the
branch in your country.
8.6
REPAIR
For all repairs before or after expiry of warranty, please return the device to your
distributor.
Voltage / current calibrator C.A 1631
31

Publicidad

loading

Este manual también es adecuado para:

Ca1631