Voltage / current calibrator C.A 1631
5 TECHNICAL CHARACTERISTICS
All of the specifications are guaranteed for one year after calibration, at
temperatures between 18°C and 28°C, with battery power.
The uncertainty is expressed as a percentage of the measurement plus the
possible variation (pts) of the least significant digit.
5.1
DC VOLTAGE INPUT AND OUTPUT
Range
100 mV
20 V
Input impedance : 2 MΩ (nominal value), < 100 pF
Over voltage protection : 30 V
Current delivered by the nominal voltage : 1 mA
5.2
DC CURRENT INPUT AND OUTPUT
Range
24 mA
Overload protection: fast-blow fuse, 125 mA,250 V
Display in percentage: 0%=4 mA,100%=20 mA
Source mode: load of 1,000 Ω at 20 mA for a battery voltage ≥6,8 V,
(700 Ω at 20mA for a battery voltage between 5.8 and 6.8V)
Simulation mode: external loop voltage condition: 24 V (nominal value), 30 V
maximum, 12 V minimum.
5.3
LOOP VOLTAGE
24 V ± 10 %
Resolution
0.01 mV
0.001 V
Resolution
0.001 mA
28
Uncertainty
± (% of the reading + pts)
0.02 % + 3
0.02 % + 3
Uncertainty
± (% of the reading + pts)
0.015 % + 3
English