22 Initial Operation

3
Initial Operation

3.1 Setting up and connecting

Setting up
Your laboratory stirrer is designed for indoor use in dry environments. When selecting a location for
your stirrer, make sure to observe the following safety instructions:
Explosion hazard! Never operate your laboratory stirrer in hazardous location!
Danger of electric shock! Never operate your stirrer in wet areas!
Fire hazard due to overheating! Never install your stirrer in furniture!
Fire hazard! For safety reasons place the stirrer at least 50 cm from any inflamma-
ble material!
Risk of tripping! Never route connection cables in highly frequented areas!
Possible cable damage! Keep away connection cable from heating zone!
When using harmful or aggressive media:
Risk of poisoning or chemical burn! The device can be damaged when sucking in
aggressive gases or vapor through the installed ventilator! laboratory stirrers must
only be operated in the presence of an exhaust system!
Install the stirrer on a flat, stable surface. The area under the stirrer must be non-combustible! Do not
put any support material under the stirrer as it blocks the ventilator installed at the bottom of the device
and may cause overheating!
Install the stirrer on a flat, clean, dry and non-slippery surface.
Avoid areas where the sun shines directly on the display (impaired readability).
Connecting
Be sure to operate the stirrer only with the specified power supply (see "2.6 Technical Data" and
details on the ratings plate).
Make sure that the mains socket is equipped with an earthing protective wire (socket outlet with
earthing)
Insert mains plug in the socket to connect the stirrer to the mains line.
Be sure to keep the socket clear for direct access in the case of emergency!
Display symbols and their meaning:
Stirrer symbol
Heater symbol
Residual-heat indicator
Bar graph
 The stirrer performs a self test which is indicated on the display
followed by information on the software version (e.g. "P 1.82").
 On completion of the self test, the display shows stirrer and
heater symbols.
 Stirrer and heater are now ready for operation.
Indicates that the stirrer is ready for operation.
Indicates that the stirrer is ready for operation.
Warning sign to inform the user that the heating zone is still hot.
Indicates the control activity of stirrer or heater.