Xy Mode; Waveform Saving Setups - Promax OS-782 Manual Del Usuario

Osciloscopio y multímetro digital
Ocultar thumbs Ver también para OS-782:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

USER'S MANUAL. OS-782

4.16.3 XY mode

This mode is only applicable to CH1 and CH2. With the XY mode selected, CH1
is displayed on the horizontal axis and CH2 is on the vertical axis; when the
oscilloscope is under the sampling mode in which no trigger is found, the data appear in
light spots.
Operations for various control keys are shown as below:
The CH1 VOL and CH1 ZORE for CH1 are used to set the horizontal
scale and position.
The CH2 VOL and CH2 ZORE for CH2 are used to set the vertical
scale and position continuously.
The following functions do not work in the XY display mode:
Reference or digital value waveform.
Cursor.
Auto Setting.
Time base control.
Trigger control.
4.17

Waveform Saving Setups

The oscilloscope can save 4 waveforms, which can be displayed on the screen
with the present waveform. The recalled waveform saved in the memory cannot be
adjusted.
The waveform saving /recalling menu is described in the following list:
Function menu
Signal source
Wave
Saving
Addresses A, B,
C, D
Page 35
Setups
Description
CH1
Select the displayed waveform which you want to
CH2
save.
MATH
A, B, C,
Select the address for saving or recalling a
D
waveform.
Store the waveform of a selected signal source into
the selected address.
Close
Close or start displaying the waveforms stored in
Start
address A, B, C or D.
02-2010

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido