10.5 Vacuum test

It is a chamber leak test or a vacuum leak test. This test aims at verifying that during the
cycle no air infiltration takes place through the chamber seals (gaskets, valves, etc...)
USE
RESPONSE
PERIODICITY
WE RECOMMEND TO USE ONLY TEST PROPOSED
34
There is no preset testing interval, apart from areas to which specific
regional regulations apply. Hence in order to constantly monitor the
autoclave performance, it is advisable to perform the test at least
once every 30 days.
Perform the test with empty chamber.
Select the suitable cycle indicated on the autoclave display. The
cycle is enabled. The autoclave automatically performs the cycle by
complying with precise technical procedures. Then the final result is
printed out through the printer connection.
The test result is immediate and is printed by the printer which prints
out all the values relating to the performed cycle as well as the final
assessment.
There is no preset testing interval, apart from areas to which specific
regional regulations apply. Hence in order to constantly monitor the
autoclave performance, it is advisable to perform the test at least
once per week.
BY TECNO - GAZ S.p.A
BEFORE THE TEST
TEST FAILED
TEST SUCCEEDED