USE AND DESTINATION OF USE OF THE AUTOCLAVE
The autoclave is able to sterilize the three types of load provided for by the standard
EN13060, especially:
METAL OR SOLID
MATERIALS
POROUS OBJECTS
HOLLOW OBJECTS
This device has been certified for the sterilization of the following materials* :
Wrapped
and
materials for a maximum load of
Type A/B hollow materials for a
maximum load of
Porous materials for a maximum
load of
* Only for European Countries
The autoclave shall only have to be used for the sterilization of tools
and materials being compatible with the steam sterilization system.
Always make sure that the loads that need to undergo sterilization
can stand the temperatures scheduled for the selected cycle.
12
Instruments with no cavities and no obstacles to the
penetration of steam
Simple or composite materials that can absorb fluids (fabrics,
gowns, surgical gauzes, dressings, etc)
Materials or devices with cavities, obstructions, etc.
These are subdivided into two types, classified according to the
length and diameter of the cavity. Approximately:
TYPE B: cannulas, tubes or devices with large passages.
TYPE A: turbines, hand pieces and devices with blind or small
holes
unwrapped
solid
ALIA B
max kg. 5
max kg. 5
max kg. 1,5
ALIA B 24
max kg. 7
max kg. 7
max kg. 2