EMV
Guidance and manufacturer's declaration – electromagnetic immunity
– for equipment and systems that are not life-supporting (IEC 60601-1-2, Table 4)
The AUTOCON
®
 III 400 is intended for use in the electromagnetic environment specified below. The customer or user of the AUTOCON
sure that it is used in such an environment.
Interference immunity tests
IEC 60601 test level
Conducted HF disturbances
3 V
effective value
acc. to IEC 61000-4-6
150 kHz to 80 MHz
Radiated HF disturbances
3 V/m
acc. to IEC 61000-4-3
80 MHz to 2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection by structures, objects
and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed HF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the AUTOCON
exceeds the applicable HF compliance level above, the AUTOCON
observed, additional measures may be necessary, such as reorienting or relocating the AUTOCON
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
EMC
Compliance level
3 V
effective value
3 V/m
®
 III 400 should be observed to verify normal operation. If abnormal performance is
CEM
®
 III 400 must make
Electromagnetic environment – guidelines
Portable and mobile HF communications equipment
should be used no closer to any part of the
AUTOCON
®
 III 400, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
f
f
for 80 MHz to 800 MHz
h
for 800 MHz to 2.5 GHz
Where P is the nominal power of the transmitter
in watts [W] according to the information provided
by the manufacturer of the transmitter and d is the
recommended separation distance in meters [m].
The field strength of stationary transmitters at all
frequencies on site
a
should be, according to a study,
less than the conformity level
b
.
Interference may occur in the vicinity of equipment
marked with the following symbol.
7
®
 III 400 is used
®
 III 400.