Cause
Lights up yellow
Significant increase in resistance
Depending on the indication, there may
be heating under the neutral electrode
Switches from green to
A significant problem occurrs when the
continuous red
unipolar current is activated.
Loosened electrode
Recognizing and remedying
errors
6. 2
Error indications for EASY
monitoring
Error indications are displayed in three stages
(green, yellow and red) when problems occur.
When working with a split neutral electrode, the
following errors may occur:
Indication
–
An acoustic signal
sounds.
An error message
appears on the display.
An acoustic signal
sounds.
A warning message
appears on the display
Detección y subsanación de
errores
Remedy
Stopping the application is not necessary.
Check the positioning of the neutral electrode.
Check the neutral electrode and neutral electrode cable,
see section EASY monitoring, page 28.
e Check the neutral electrode cable for proper connection
and external damage.
e Correct the positioning of the neutral electrode. If the
error persists, replace the neutral electrode.