Configuration - Dräger X-am 2500 Instrucciones De Uso

Ocultar thumbs Ver también para X-am 2500:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 57
WARNING
!
CSA requirement: Measured values over the full scale value
may indicate an explosive atmosphere.
Only applies for class II certification. CSA standard C22.2 no.
152 does not contain any requirements for class II hazard
areas. Therefore this device was not tested for class II by the
CSA. The sensor may become blocked and not measure the
gas correctly or not warn the user that gas measurement is
not possible.
WARNING
!
The sensitivity must be tested every day before first use using a
known concentration of the gas being measured corresponding
to 25 to 50 % of the full concentration value. The accuracy must
be 0 to +20 % of the daily value. The accuracy may be corrected
via calibration.
NOTICE
i
i
CSA requirement: Only the combustible gas detection portion
of this instrument has been assessed for performance.
The instrument has not been classified by the CSA for use in
mines.
Dräger X-am 2500
2.3
Approvals
The approvals are shown on the rating plate, see "Notes on Approval"
on page 417. Do not stick anything on the name plate on the gas
detector. The technical suitability tests are valid for the X-am 2500 gas
detection instrument and the calibration cradle. The explosion-
protection approvals are only valid for the X-am 2500 gas detection
instrument; the calibration cradle must not be used in the Ex zone.
The BVS 10 ATEX E 080 X technical suitability test is based on the
adjustment with the target gas.
.
CE marking:
See declaration of conformity on page 418.
3

Configuration

NOTICE
i
i
Only trained persons are permitted to carry out modifications
to the instrument configuration.
To individually configure an instrument with standard configuration,
the instrument must be connected to a PC via the USB infrared cable
(order no. 83 17 409). The configuration work is carried out using the
PC software Dräger CC-Vision. The PC software Dräger CC Vision
can be downloaded from the following web address free of charge:
www.draeger.com/software.
Change configuration: see Technical Manual.
Configuration
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mqg 0011

Tabla de contenido