Ivoclar Vivadent Bluephase Style M8 Instrucciones De Uso página 75

Ocultar thumbs Ver también para Bluephase Style M8:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50
Guidance and manufacturer's declaration - electromagnetic immunity
The Bluephase Style M8 is intended for use in the electromagnetic environment specified below. The customer or the user of the Bluephase Style M8
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 V
rms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broad-
cast cannot be predicted theoretically with accuracy. To assess the electromagnetic environ-
Recommended separation distances between portable and mobile RF communications equipment
and the "Bluephase Style M8"
The Bluephase Style M8 is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the Bluephase Style M8 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Bluephase Style M8 as recommended below, according to the maximum output power of the
communication equipment.
Separation distance according to frequency
Rated maximum
output power
of transmitter
150 kHz to
W
80 MHz
d = 0.35 √P
0.01
0.04
0.1
0.13
1
0.40
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Compliance level
10 V
10 V/m
of transmitter
m
80 MHz to
800 MHz
d = 0.35 √P
0.04
0.13
0.40
1.3
1.3
4.0
4.0
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should not be used
closer to any part of the Bluephase Style M8, including cables, than the
recommended separation distance caculated from the equation applica-
ble to the frequency of the transmitter.
Recommended separation distance
d = 0.35 √P
d = 0.35 √P 80 MHz to 800 MHz
d = 0.70 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the trans mitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electro-
magnetic site survey,
frequency range.
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
ment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Bluephase Style M8 is used exceeds the
applicable RF compliance level above, the Bluephase Style M8 should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be neces-
sary, such as reorienting or relocating the Bluephase Style M8.
b
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.
800 MHz to
2.5 GHz
d = 0.7 √P
0.07
0.22
0.70
2.2
7.0
a
should be less than the compliance level in each
75

Publicidad

Tabla de contenido
loading

Tabla de contenido