Heat development
As it is the case with all high-performance lights, the
high light intensity results in a certain heat develop-
ment. Prolonged exposure of areas near the pulp and
soft tissues may result in irreversible damage. Therefore,
the recommended curing times have to be observed,
especially in areas near the pulp (adhesives 10 seconds).
Uninterrupted curing times of more than 40 seconds on
the same tooth surface, as well as direct contact with the
gingiva, oral mucous membrane or skin have to be pre-
vented. Polymerize indirect restorations at intermittent
intervals of 20 seconds each or use external cooling with
an air stream.
3. Start-up
Check the delivery for completeness and any possible
transportation damage (see delivery form). If parts are
damaged or missing, contact your Ivoclar Vivadent rep-
resentative.
Charging base
Before you switch on the device, make sure that the volt-
age mentioned on the rating plate complies with your
local power supply.
Connect the power cord with the power supply and the
power pack. The Power On indicator on the power pack
lights up in green and the illuminated ring on the charg-
ing base flashes briefly.
7