Handpiece
Remove the handpiece from its packaging and clean the
light probe and the handpiece (see chapter Maintenance
and cleaning). Slightly turn the light probe to remove
and reposition it.
After that, mount the anti-glare cone on the light probe.
Battery
Fully charge the battery before the first use! If the bat-
tery is fully charged, it features a curing capacity of
approximately 20 minutes. Slide the battery straight into
the handpiece until you hear and feel it click into place.
8
Gently place the handpiece into
the corresponding rest on the
charging base. If a hygiene sleeve
is used, please remove it before
you charge the battery. If possible,
use the light always with a fully
charged battery. This will prolong
the service life. It is therefore rec-
ommended to place the handpiece
into the charging base after each
patient. If the battery is fully dis-
charged, the charging time is
2 hours.
Since the battery is a consumable, it has to be replaced
after its typical life cycle has expired after approximately
2.5 years. See battery label for the age of the battery.
Battery charging status
The respective charging status is indicated on the
handpiece as described on Page 4.