Operation
Disinfect contaminated surfaces of the curing light as
well as light probes and anti-glare cones before each
use. Furthermore, make sure that the stipulated light
intensity permits adequate polymerization. For that
p urpose, check the light probe for contamination and
damage as well as the light intensity at regular intervals.
Selecting the curing time
The curing time can be individually set using the time
selection button. Users may choose between 10, 15, 20
and 30 seconds.
Observe the Instructions for Use of the material applied
when selecting the curing time. The curing recommenda-
tions for composite materials apply to all shades and, if
not mentioned otherwise in the Instructions for Use, to a
maximum layer thickness of 2 mm. Generally, these
Curing times
Restorative materials
Indirect restorations /
cementation materials
Adhesives
Temporary materials
Miscellaneous
1) Applies to a maximum layer thickness of 2 mm and provided that the Instructions for Use of
the respective material do not state any other recommendation (might be the case e.g. with
dentin shades)
2) Applies to a maximum layer thickness of 4 mm and provided that the Instructions for Use of
the respective material do not state any other recommendation (might be the case e.g. with
dentin shades)
Composites
• 2 mm
1)
Tetric EvoCeram / Tetric EvoFlow / IPS Empress Direct /
Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
All conventional composites
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Compomers
3)
Compoglass F / Compoglass Flow
Variolink Base
4)
/ Variolink Veneer /
Variolink N Base
/ Variolink N Clear Veneer
4)
Variolink Ultra
5)
Dual Cement
/ Variolink II
5)
5)
Syntac / Heliobond / ExciTE F / ExciTE F DSC / AdheSE / AdheSE One F /
Tetric N-Bond / Tetric N-Bond Self-Etch
Systemp.inlay/onlay / Fermit / Fermit N
Telio CS Link / Systemp.link
Helioseal / Helioseal F / Helioseal Clear
Monopaque
MultiCore Flow / MultiCore HB
IPS Empress Direct Opaque
IPS Empress Direct Color
Telio Add-On Flow
Telio Stains
Heliosit Orthodontic
Vivaglass Liner
re c ommendations apply to situations where the emission
window of the light probe is placed directly over the
material to be polymerized. Increasing the distance
between the light source and the material will require
the curing time to be extended accordingly. For instance,
if the distance to the material is 9 mm, the effective light
output is reduced by approx. 50%. In this case, the
w recommended curing time has to be doubled.
3) Applies to a maximum layer thickness of 3 mm
4) Applies to light-curing (use of the base past only)
5) Applies to dual-curing
Start button
Time selection
button
1,100 mW/cm
2
± 10%
10 s
15 s
10 s
20 s
per mm ceramic:
10 s per surface
per mm ceramic:
10 s per surface
10 s
10 s
10 s per surface
10 s
20 s
15 s
10 s
11