Operation
Disinfect or autoclave contaminated surfaces of the curing
light as well as of the light probes and anti-glare cones
before each use. Furthermore, make sure that the stipulated
light intensity permits adequate polymerization. For this
purpose, check the light probe for contamination and dam-
age as well as the light intensity at regular intervals (e.g.
with the Ivoclar Vivadent radiometer "bluephase meter").
Selecting the curing program and the
curing time
The curing program and the curing time can be individually
set. bluephase C8 is equipped with the following 3 curing
programs for the different indications. The desired curing
program is selected with the program selection buttons. The
display changes accordingly (see Indicators on the hand-
piece). The device comes equipped with the following pre-
set programs:
Factory settings
HIGH
(High Power)
LOW
(Low Power)
SOFT
(Soft start)
12
15 seconds
10 seconds
20 seconds
The intended curing time is selected
using the time selection buttons. Users
may choose between 5, 10, 15, 20 and
30 seconds.
Observe the Instructions for Use of the material applied
when selecting the curing time.
The curing recommendations for composite materials apply
to all shades and, if not mentioned otherwise in the Instruc-
tions for Use, to a maximum layer thickness of 2 mm. Gen-
erally, these recommendations apply to situations where the
emission window of the light probe is placed directly over
the material to be polymerized. Increasing the distance
between the light source and the material will require the
curing time to be extended accordingly. For instance, if the
distance to the material is 9 mm, the effective light output
is reduced by approx. 50%. In this case, the recommended
curing time has to be doubled.