Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
Table of Contents

Specifications

Specification
Details
Size range
0.3 µm - 10 µm channels per ISO 14644-1 (FS 209E)
industry standard; HHPC 3+ (0.5 µm) and HHPC 2+:
0.5 µm-5 µm
Number of channels
HHPC 6+: Channel 1 fixed at 0.3 µm, and up to five
additional channels with sizes selectable from 0.5, 1.0,
2.0, 5.0 or 10.0 µm.
HHPC 3+ (0.3 µm): Channel 1 fixed at 0.3 µm, and up
to two additional channels with sizes selectable from
0.5, 1.0, 2.0, 5.0 or 10.0 µm.
HHPC 3+ (0.5 µm): Channel 1 fixed at 0.5 µm, and up
to two additional channels with sizes selectable from
1.0, 2.0 or 5.0 µm.
HHPC 2+: Channel 1 fixed at 0.5 µm, and up to one
additional channel with sizes selectable from 1.0,
2.0 or 5.0 µm.
Weight
0.7 kg (1.5 lbs)
Flow rate
2.83 lpm. Get ISO 14644-1 class 5 compliant 0.3 µm
samples in a minute.
USB support
Simple data export to memory stick or direct connect to
PC via USB cable. No software required.
Specification
Details
Ethernet support
Access particle data over network with an internet
browser such as Windows Explorer, Safari and Firefox
Display
3.5" High resolution 320 x 240 color
Display configuration
User selectable size channels
Display data modes
Traditional tabular particle counts or trend graph
Enclosure
High-impact polycarbonate ABS
Enclosure rating
IP40 Environmental rating. Protected from objects
1 mm in diameter or larger. Not protected from water.
Dimensions (L x W x D) 272 x 99 x 54 mm (10.7 x 3.9 x 2.1 in.)
Battery duration
More than 10 hours with typical use model. Minimum
5.5 hours continuous sampling.
Battery charging
3.5 hours
Data storage
10,000 records (5,000 records for .xls CFR 21 Part
11 secure files)
Counting efficiency
50% at 0.3 µm; 100% for particles > 0.45 µm (per ISO
21501)
Zero count level
1 count every 5 minutes (per JIS B9921)
Concentration limits
10% at 4,000,000 per cubic foot (per ISO21501)
Count modes
Raw counts, N/CF, N/CM, N/L in Cumulative or
Differential mode
Security
Administrator password controlled (optional)
UI Languages
English and Japanese
Alarms
User-selected particle channels and limits
Operating environment
10 ºC to 40 ºC (50 ºF to 104 ºF) / < 95% non
condensing. Pollution degree: II or better.
Storage environment
-10 ºC to 50 ºC (14 ºF to 122 ºF) / Up to 98% non-
condensing
English 3

Publicidad

Capítulos

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Hhpc 3+Hhpc 2+

Tabla de contenido