ROHO DRY FLOATATION Wheelchair Cushions Operation Manual
Cushion and Cover Care
Warnings:
- Cleaning and disinfecting are separate processes. Cleaning must precede disinfection. Before use by a
different individual: clean, disinfect, and check the product for proper functioning.
- Follow all manufacturer bleach container and germicidal disinfectant safety guidelines.
- Note about sterilization: High temperatures accelerate aging and will damage the product assembly.
The products featured in this manual
or required to be sterilized prior to use. If institution protocol demands sterilization: First, follow
the cleaning and disinfecting instructions; then, open the inflation valve and use the lowest possible
sterilizing temperature, but not to exceed 70°C (158°F), for the shortest time possible. DO NOT
autoclave.
Cover Care
Remove hook
fastener pieces from the bottom of the cover prior to cleaning and disinfecting. Wash with like
colors.
Clean the Cover: Remove the cover. Machine wash in cold water (30°C) with mild detergent, gentle cycle, or
wipe clean with neutral detergent and warm water. Tumble dry low.
Disinfect the Cover: Hand wash in cold water (30°C) using 1 part household liquid bleach per 9 parts water.
Keep the product wet with the bleach solution for 10 minutes, then rinse thoroughly. Or machine wash in hot
water (60°C). Tumble dry low.
Cushion Care
Warning: Inflation valve(s) MUST be closed to make sure that water does not enter the product during
cleaning or disinfecting.
Caution: Use of the following may cause damage to the cushion: abrasives (steel wool, scouring pads);
caustic, automatic dishwasher detergents; cleaning products that contain petroleum or organic solvents,
including acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based
lotions, lanolin, ozone gas; ultraviolet light cleaning methods.
Caution: DO NOT expose to direct sunlight.
Hand Wash and Hand Disinfect the Cushion (Preferred Method)
Hand Wash the Cushion: Remove the cover, close the inflation valve(s) and place in a large sink. Hand wash,
using mild liquid hand soap, hand dishwashing liquid, laundry detergent, or a multipurpose
detergent (Follow product label instructions.) Gently scrub all cushion surfaces, using a soft
plastic bristle brush, a sponge or a cloth. Rinse with clean water. Air dry.
Caution: Rinse thoroughly. Detergent residue may cause air cells to stick together.
Hand Disinfect the Cushion: Follow the hand wash instructions, using 1 part household
liquid bleach per 9 parts water. Keep the product wet with the bleach solution for 10
minutes. Rinse thoroughly with clean water. Air dry.
are not packaged in a sterile condition, nor are they intended
9
30C
or
+
or
30C
1: 9
+