3.5 Control Box

The control box is latched to the backrest motor.
The control box A is provided with a locking cam B and a label with
symbols showing where to connect the motor plugs:
Backrest motor C
Thigh section's motor D
Bed end motor, foot end E
Bed end motor, head end F
Hand control G

3.6 Wiring

The control box is placed underneath the upper half of
the mattress support.
The control box is provided with labels showing where
to connect the different motor wires.
WARNING!
Wires damaged or torn apart could lead to
personal injuries or death.
Follow instructions for wiring carefully to make sure
there is no squeezing risk.
Before the bed is put into use:
– Check that the wires are attached to the top frame
according to instructions.
– Check that the wires are kept clear of the floor and
do not block the castors.
– Check the functions by operating the motors of the
bed to their outer positions.
– Check that the wires are not being squeezed during
operation of the functions.
IMPORTANT!
– The wires must be mounted in such a way that they are
kept clear of the floor and do not block the castors.
It is normal for wires to loosen slightly after short usage.
IMPORTANT!
The hand control, control unit and motors are protected
according to IPX4.
– A lock cam must be used on the control box, if
Invacare® is to guarantee the IP protection.
8
1.
Connect both plugs from the hi/lo motors (head end and foot
end) to the control box placed underneath the backrest.
2.
Guide the hi/lo motor wire from the bed foot end AND the wire
from the thigh/leg section motor through the thigh/leg section
motor pipe pin. Make sure the pipe pin is properly locked.
3.
Connect the control box to the mains.
4.
Run the thigh/leg section to its highest positions .
5.
Place the wire hi/lo motor in the foot end on the hook at the
bed end.
6.
Run the backrest to its highest position
7.
Place the wire from the hi/lo motor in the head end, on the two
hooks at the bed end and attach the wire in the backrest motor
pipe pin.
8.
Install the safety clip over the plugs in the control box.