Connect To A Flow Logger Or Sampler; Install The Sensor With A Primary Device - Hach US9001 Manual Del Usuario

Ocultar thumbs Ver también para US9001:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 63
Figure 7 Mounting hardware options – US9001
Figure 8 Mounting hardware options – US9003

Connect to a flow logger or sampler

Connect the cable to a flow logger or sampler. Refer to the flow logger or sampler documentation for
instructions.

Install the sensor with a primary device

A primary device is a hydraulic structure, such as a flume or weir, that has a known level-to-flow
relationship. The ultrasonic sensor measures the level of liquid in a channel that contributes to flow
(referred to as the "head") and the flow logger calculates the flow rate based on the head-to-flow
relationship of the primary device.
Note: As an alternative, the flow logger can calculate the flow rate based on the level of liquid in the channel and
the area velocity supplied by a velocity sensor. The flow rate calculation method is selected in the Flow Method
field.
10 English

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Us9003

Tabla de contenido