5.3 IRREGULAR HEARTBEAT DETECTOR

An irregular heartbeat is detected when a heartbeat rhythm varies
while the device is measuring systolic pressure and diastolic pressu-
re. During each measurement, blood pressure monitor will keep a
record of all the pulse intervals and calculate the average value of
them. If there are two or more pulse intervals , the difference betwe-
en each interval and the average is more than the average value
of ±25% , or there are four or more pulse intervals ,the difference
between each interval and the average is more than the average
value of ±15%, then the irregular heartbeat symbol will appear on
the display with the measurement result.
5.4 WHY DOES MY BLOOD PRESSURE FLUCTUATE THROUGHOUT
THE DAY?
1.
Individual blood pressure varies multiple times everyday. It is
also affected by the way you tie your cuff and your measu-
rement position, so please take the measurement under the
same conditions.
2.
If the person takes medicine, the pressure will vary more.
3.
Wait at least 3 minutes for another measurement.
6. ERROR MESSAGE MALFUNCTION
PROBLEM SYMPTOM
Display is dim or
No power
will not light up.
Low
Show on
batteries
the display
E1 shows
E 2 shows
E 3 shows
E 10 or
E 11 shows
Error
E 20 shows
message
E 21 shows
EExx,shows on
the display.
Warning
"out " shows
message
CHECK THIS
Batteries are exhausted.
Batteries are inserted
incorrectly.
Batteries are low.
The cuff is not secure.
The cuff is very tight
The pressure of the
cuff is excess.
The monitor detected motion,
talking,or the pulse is too poor
while measuring.
The measurement process
does not detect the pulse
signal.
Measure incorrectly .
A calibration erro
r
occurred.
Out of measurement
range
REMEDY
Replace with new batteries
Insert the batteries correctly
Replace with new batteries
Refasten the cuff and then measure again.
Refasten the cuff and then measure again.
Relax for a moment and then measure again.
movement can affect the measurement.
Relax for a moment and then measure
again.
Loosen the clothing on the arm and then
measure again.
Relax for a moment and then measure again.
Retake the measurement.If the problem
persists,contact the retailer or our customer
service department for further assistance.
Refer to the warranty for contact information
and return instructions.
Relax for a moment. Refasten the cuff and
thenmeasure again. If the problem persists,
contact your physician.
EN