Optional Automatic Wash:
Any deviation from this guideline could result in damage to equipment as well as
improper cleaning results. Parameters given below must be followed exactly in order
to assure proper cleaning and avoid damage to the equipment.
• Disassemble the device as indicated in section Cleaning of the MAYFIELD Infinity XR2 Skull
Clamp.
• Rinse components under warm tap water prior to placing in washer.
• Arrange device in a way to avoid contact between components.
• Select instrument cycle and ensure proper programming.
Table 3 Instrument Cycle
Phase
Pre-Wash 1
Enzyme Wash
Wash 1
Rinse 1
Thermal Rinse**
* Hot Tap water not to exceed 82.2° C
** Optional phase for disinfection of components
• Remove from washer and dry completely if needed.
• Inspect all components to ensure there is no visible organic debris or residue from cleaning
agent. Repeat process if any visible soil is detected.
(For additionally validated cleaning cycles, refer to Appendix A)
* No limit on tap water temperature
** Optional phase for disinfection of components - minimum water temperature as indicated or
per washer manufacturer specifications for the thermal rinse cycle
CAUTION!
Time in minutes
04:00
04:00
10:00
00:30
02:00
Table 3 Instrument Cycle
Page 25
Water
Detergent type &
Temperature
Concentration
Cold tap H
O
2
Endozime® AW
Hot tap H
O*
Triple plus w/
2
A.P.A., 1:128 ratio
Renu-Klenz™,
60° C
Hot tap H
O*
2
82.2° C
N/A
1:256 ratio
N/A
N/A