Bleach Dilution:
Caution: Bleach solutions produce chlorine gas. Prepare in a well ventilated area.
NOTE: Hypochlorite potency decreases as bleach ages. Use only recently purchased bleach to prepare
decontamination solutions. Do not use pre-diluted bleach. Prepare a fresh batch for each application.
Hypochlorite bleach mixing guidelines to obtain 20,000 ppm solution:
NOTE: The use of highly alkaline agents may cause some components to tarnish and/or corrode. While
this does not affect the performance of the device we recommend that you use an acidic neutralizing
agent immediately after each use as this will help keep your device looking like new.

Maintenance and Care

The Infinity XR2 Skull Clamp should be returned to its storage area after each use whether it be a
designated OR or equipment room.
NOTE: Ensure all components are completely dried after each cleaning.
If the Skull Clamp is dropped or mishandled, it should be inspected for damage. (REF
Inspection section of this Instruction manual). If damage occurs, do NOT use; return
the complete device immediately to Integra LifeSciences for inspection.
NOTE: Follow hospital procedures for disposal of this device.
Bleach
Concentration
4.5%
5%
6%
Table 5 Hypochlorite Bleach Mixing Guidelines.
CAUTION!
Preparation Instructions
Pour 4 units bleach into 5 units water
Pour 2 units bleach into 3 units water
Pour 1 unit bleach into 2 units water
CAUTION!