Cleaning

NOTICE
Do not use soapy water to clean the dosing unit.
After cleaning the measurement device, ensure that each dosing unit is
screwed back into the corresponding connection on the measurement
device.
Clean the device with compressed air.
Clean out the Dräger tube holder (rubber part) with compressed air af-
ter each use.
If the measurement has detected that the measured air is contaminated:
Clean the device with clean air for at least 6 minutes.
Dismantle the measurement device in cases of greater contamination,
and clean the individual parts (except the dosing unit) with soapy water.
Dry with oil-free, dry air.
Only clean the dosing unit with
oil-free, dry air in the direction
indicated by the arrow.
Assemble the measuring de-
vice.
If the measurement has detected
that the air is contaminated with oil (only applies for high-pressure
systems):
Take out the pressure reduc-
er's air filter.
Insert a new air filter and screw
into place.
Clean the device with clean air
for at least 6 minutes.
Visit: www.thesafetyequipmentstore.com or email besafe@thesafetyequipmentstore.com for sales and service.
32
Dismantle the measurement device in cases of greater contamination,
and clean the individual parts (except the dosing unit) with soapy water.
Dry with oil-free, dry air.
Only clean the dosing unit with
oil-free, dry air in the direction
indicated by the arrow.
Assemble the measuring
device.

Storage

Store all parts without deformations in a dry, cool, dust-free location. Keep
out of direct sunlight and heat. Strictly follow ISO 2230.
Store Dräger tubes safely away from unauthorised persons.