Order List; Cleaning And Disposal; Technical Data - Dräger Infinity ID Instrucciones De Uso

Circuito respiratorio calefactado(n)
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 8
CAUTION
Risk of burns
The heated breathing circuit must not be covered with bed sheets, hand
towels or similar flammable objects.
WARNING
Fill the humidifier chamber with distilled water before using the
breathing circuit and check the function of the floater.
WARNING
Risk of patient injury
Only use distilled water. The use of other substances can harm
the patient.
WARNING
Risk of patient injury
The port at the Y-piece is only for pressure measurement. Do not
use for the temperature sensor.
CAUTION
Risk of patient injury
Too high or too low ambient temperatures can lead to increased accu-
mulation of humidity or humidification performance can be reduced.
Do not operate the device outside the stated ambient temperature
range.
CAUTION
Risk of aspiration of condensate
When the breathing circuit with hose extension is used outside of the
incubator, more condensate accumulates.
Dismantle the hose extension when using the breathing circuit outside
of the incubator.
NOTE
This medical device has been designed, tested, and manufactured
exclusively for single-patient use and for a period of use not exceeding
7 days.

Cleaning and disposal

The user is responsible for regularly replacing the medical device
according to the hygiene regulations.
WARNING
Reuse, reprocessing, or sterilization can lead to failure of the
medical device and cause injury to the patient.
This medical device has been designed, tested, and
manufactured exclusively for single use. The medical device
must not be reused, reprocessed, or sterilized.
Following use, the medical device must be disposed of according to local
public health and waste disposal regulations in order to avoid possible
contamination.

Technical data

Breathing hose length
1.2 m (47 in)
Material
Breathing hoses
PP
Connectors
PP, SBC, TPR
Y-piece
PC
Humidifier chamber
PC, SBC, silicone, aluminum
All gas-conducting components are free of PVC.
Performance data
Breathing circuit
Insp./exp. resistance at 60 L/min
<15.2 mbar
<12 mbar (<12.3 cmH
Insp./exp. resistance at 30 L/min
<7 mbar (<7 cmH
<4 mbar (<4 cmH
Insp./exp. resistance at 15 L/min
<2 mbar (<2 cmH
<1.1 mbar (<1.1 cmH
Insp./exp. resistance at 5 L/min
<0.4 mbar (<0.4 cmH
<0.2 mbar (<0.2 cmH
Insp./exp. resistance at 2.5 L/min
<0.1 mbar (<0.1 cmH
<0.1 mbar (<0.1 cmH
Compliance at 60 mbar
<0.9 mL/mbar (<0.9 mL/cmH
Leakage at 60 mbar
<50 mL/min
Maximum operating pressure
60 mbar
Electrical connection data
22 V, 2.73 A, 60 W, 16
Humidifier chamber volume
130 mL
Inspiratory limb volume
500 mL
Warm-up time
30 min
Humidifier output:
Invasive ventilation at 4 to 30 L/min >33 mg/L
Non-invasive ventilation at 4 to
>10 mg/L
30 L/min
Ambient conditions
During storage
Temperature
–20 °C to 60 °C (–4 °F to 140 °F)
Relative humidity
5 % to 95 % (non-condensing)
Ambient pressure
500 hPa to 1200 hPa
(7.3 psi to 17.4 psi)
During operation
Temperature
18 °C to 26 °C (64 °F to 79 °F)
Relative humidity
5 % to 95 % (non-condensing)
Ambient pressure
500 hPa to 1200 hPa
(7.3 psi to 17.4 psi)
Instructions for use Infinity ID Breathing Circuit heated (N)
Classification according to
Directive 93/42/EEC Annex IX
UMDNS code
Universal Medical Device
Nomenclature System –
nomenclature for medical devices
Protection class
1)
1 bar = 1 kPa x 100

Order list

Designation
Infinity ID Breathing Circuit heated (N)
Softconnector, pack of 10
Additional items according to Dräger accessories
catalog
1)
(<15.5 cmH
O) /
2
O)
2
O) /
2
O)
2
O) /
2
O)
2
O) /
2
O)
2
O) /
2
O)
2
O)
2
Class IIa
14-238
Type BF
Order No.
MP00343
MP01339
English/English US
5