Safety
6
Medial support joint with dummy in 5° or 7.5° (not illus­
trated)
INFORMATION
The subsequent conversion of the flexion components to 7.5° may only be completed by an authorised Ottobock
Service Centre.
1.6.5 Construction
The knee joint has a 6 mm posterior placement. This facilitates easy extension of the knee joint, which assures
problem-free unlocking.
The joint bushing (radial bearing) and bearing washer (axial bearing) are made of maintenance-free plastic. The
wedge lock is connected to an electronically controlled spring, which locks it to the lower joint section. The elec­
tronic control unit consists of a gyroscope and 2D acceleration sensors to record the gait phase. It measures the
position while walking and releases the lock prior to toe-off while locking it again in the extended position prior to
heel strike.
The E-MAG Active has an additional locking function (PreLock) that is activated at 15° of flexion. This secures the
patient even before the swing phase as such ends. Subsequently the knee joint is fully locked when it reaches the
extension stop.
The open joint bar cases serve for attachment of the lamination bars. The electronic components are designed so
that they can be replaced by a technician. The batteries can be removed from the lamination frame and charged in
the charger. The charger has two ports so that two batteries can be charged simultaneously.
1.6.6 Support
Your national Ottobock team will be happy to answer any technical questions. Contact addresses and telephone
numbers are listed on the last page of this manual.
2 Safety
2.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
2.2 General safety instructions
INFORMATION
Certification (certificate of competence) for the E-MAG Active is mandatory. Without an orthopaedic technician
certified by name, the E-MAG Active is not CE compliant and therefore CE certification expires from an industrial
perspective.
CAUTION
Risk of falling due to improper use
Provide your patient with precise instructions on the operation and use of the E-MAG Active. Be sure to inform
your patient that, in case of improper use, the joint cannot release prior to toe-off or the swing phase, or cannot
lock prior to heel strike. Also inform your patient of the PreLock function.