3.5. POWER SUPPLY

3.5.1. J93 CLAMP
The power supply of the J93 clamp is provided by a 9 V alkaline battery (type 6LF22, 6LR61, or NEDA 1604).
The nominal operating voltage is between 6.5 and 10 V.
The battery life is 70 hours. The typical consumption is 11 mA.
During the battery test, if the indicator lights, the clamp can be used.
When the indicator no longer lights, the battery must be replaced (see §4.2).
3.5.2. J193 CLAMP
The J193 clamp is supplied by the device to which it is connected. It delivers a voltage of 5 V ±0.25 V.
The typical consumption of the clamp is 5.4 mA.

3.6. ENVIRONMENTAL CONDITIONS

The device must be used under the following conditions:
% RH
100
95
90
80
75
70
60
3
50
40
30
20
10
0
-50 -40 -30 -20 -10 0
For an extended period of non‑use or storage, remove the battery from the J93 clamp.
Indoor use.
Degree of pollution: 2.
Altitude: <2000m.
1
2
18
28 35
55
10 20 30 40 50 60 70 80 90
1: Reference range
2: Operating range
3: Storage range (without battery)
°C
26