GB | INSTRUCTIONS FOR USE
Manual cleaning/disinfection
Wipe disinfection for electrical devices without
sterilization
Phase
I
Step
Wipe disinfection
T (°C/°F)
RT (room temperature)
≥1
t (min)
Conc. (%)
-
Water quality
-
Chemical
Meliseptol HBV wipes 50%
Propan-1-ol
Phase I
Remove any visible residue with a dis-
•
posable disinfectant wipe.
Wipe all surfaces of the visually clean
•
product with a fresh disposable disinfec-
tant wipe.
Observe the specified contact time (1
•
minute minimum).
Inspection
Check product for damages after each
•
cleaning/disinfection.
Immediately set aside product if dam-
•
aged.
Storage
Store products protected from dust in
•
a dry, dark and cool room with as little
bacterial contamination as possible.
42
Reader Unit Set