Reader Unit Set
TAKING MEASUREMENTS
Press the <Menu> button to open the [Menu].
The following information is displayed on the
screen:
The <OK> button to open the [Measurement]
submenu. The following information is dis-
played on the screen:
There are three types of measurement:
1. [Single measurement]: The mean of
8-10 measurements is displayed and
saved to the SD card.
2. [Continuous measurement]: The
sequence of single measurements within
the selected time interval is shown as a
time diagram and save to the SD card.
3. [Fast measurement]: The sequence of
unaveraged single measurements at the
maximum available measurement rate
(max. 44 measurements per second) is
shown as a time diagram and save to
the SD card.
INSTRUCTIONS FOR USE | GB
Single measurement
Press the <OK> button to select the [Single
measurement] menu.
The following information is displayed on the
screen:
Press the <Start> button to start the [Single
measurement].
The following screen content is displayed:
Continuous measurement
In the [Measurement] menu, press the <Up
arrow> or <Down arrow> button to select the
[Continuous measurement] menu and confirm
with <OK>. The following information is dis-
played on the screen:
Press the <Start> button to start the [Continu-
ous measurement].
The following information is displayed on the
screen:
Press the <Star> button to place a marker.
Multiple markers can be placed during a mea-
surement process. Press the <Stop> button to
stop the measurement.
Markers permit the circumstantial analysis of
the measured data.
Symbols for reception quality
Symbol
Communication started
Distance from antenna
to measuring cell: - good
Distance from antenna
to measuring cell too
close: - increase distance
Distance from antenna
to measuring cell too
large: - decrease distance
Fast measurement
There are two ways to start a [Fast measure-
ment]:
On the start screen, press the <Fast mea-
surement> button directly and then start the
measurement with the <Start> button. Alter-
natively, you can press the <Menu> button to
display the following submenu:
Press the <OK> button to open the [Mea-
surement] menu. In the [Measurement] menu,
Explanation
37