Reader Unit Set
Message displayed
SD card not readable!
Info> button: Measurement with-
out data storage possible!
SD card memory full. Measure-
ment without data storage possi-
ble!
SD card memory full
SD card memory almost full
System error
<Info> button: Ambient pressure
not readable
System error - incompatibility
System error - antenna incompat-
ible
System error - ID data inadmissi-
ble
System error - implant voltage out
of range
System error - calibration data
incorrect
System error - contact Technical
Service
Keypad faulty
Cause
SD card missing - or - SD card ID
not readable - or - SD card con-
tains no calibration data
SD card memory full (100%) dur-
ing continuous measurement
SD card memory full (100%) dur-
ing continuous measurement
SD card memory almost full
(99%) during continuous or fast
measurement
Barometric pressure (BARD) not
readable
Device hardware and software
versions not compatible
Hardware version of antenna and
Reader Unit not compatible
Implant identification data corrupt
Implant voltage outside of permis-
sible range
Calibration data in implant cor-
rupted or not readable (only
applies if no SD card has been
inserted)
System test detected an error
A pressed button is registered
when switching on the Reader
Unit
INSTRUCTIONS FOR USE | GB
Error detection / troubleshooting
Insert the matching M.scio SD card
into the Reader Unit. Measurement
with data storage possible, but with
limited accuracy.
Delete any data no longer required,
measurement without data storage
possible.
Delete any data no longer required.
Delete any data no longer required.
Device function locked. Deactivate
and reactivate device. If the error
persists, contact Technical Service.
Device function locked. Deactivate
and reactivate device. If the error
persists, contact Technical Service.
Device function locked. Deactivate
and reactivate device. If the error
persists, replace antenna and/or
contact Technical Service.
Measurement is stopped. Deacti-
vate and reactivate device. If the
error persists, contact Technical
Service.
Measurement is stopped. Deacti-
vate and reactivate device. If the
error persists, contact Technical
Service.
Measurement is stopped. Deac-
tivate and reactivate device. If
the error persists, contact Tech-
nical Service. Measurement with
inserted SD card possible.
Device function locked. Deactivate
and reactivate device. If the error
persists, contact Technical Service.
Release the key. Deactivate and
reactivate device.
47