Reader Unit Set
ELECTROMAGNETIC EMISSIONS
CAUTION
The device is only intended for use at professional healthcare facilities!
While the device complies with class B HF emissions in accordance with CISPR 11, it may only be
used at professional healthcare facilities.
Guidelines and manufacturer's declaration - electromagnetic emissions
The device is intended for use in an electromagnetic environment as specified below. It is the
user's responsibility to ensure that the device is operated in such an environment.
Electromagnetic interfer-
ence measurements
RF emissions according
to CISPR 11
RF emissions according
to CISPR 11
Harmonic emissions
according to IEC
61000-3-2
Voltage fluctuations/
flicker emissions accord-
ing to IEC 61000-3-3
ELECTROMAGNETIC IMMUNITY
WARNING
Creation of electromagnetic interferences!
Portable HF communication devices (radio devices) (including their accessories, such as antenna
cables and external antennas) should not be used closer than a distance of 30 cm (or 12 inches) to
the reader and the antenna. If electromagnetic interferences with spot frequencies of 385 MHz or 450
MHz arise, a distance of at least 80 cm must be ensured. Non-compliance may lead to a reduction
of the device's performance characteristics. Electromagnetic interferences can lead to a device shut-
down. In this case, the device must be restarted and the measurement repeated.
MEDICAL DEVICES CONSULTANTS
In
compliance
with
on medical devices (directive 93/42/EEC),
Christoph Miethke GmbH & Co. KG has nom-
inated medical product consultants as con-
tacts for all product-related questions.
You
can
reach
consultants at:
Phone +49 331 62083-0
info@miethke.com
Compliance
Complies with
Group 1
Complies with
Class B
Complies with
Class A
(IEC 61000-3-2)
Complies
European
directive
our
medical
devices
INSTRUCTIONS FOR USE | GB
Guidelines on the electromagnetic environ-
ment
The device uses RF energy only for its inter-
nal function. Its RF emissions are thus very
low and not likely to cause interference in
nearby electronic devices.
The device is suitable for use in all establish-
ments including domestic settings and areas
immediately connected to a public power
supply which also supplies buildings used for
residential purposes.
-
-
51