GB | INSTRUCTIONS FOR USE
Message displayed
Wrong SD card inserted! Remove
card!
<Info> button: Measurement with-
out data storage possible - or -
Insert SD card with correct ID
XXXXXXXXXX!
Wrong implant - restart measure-
ment!
<Info> button: Switch to another
implant during continuous mea-
surement not possible!
Device temperature outside of
range
<Info> button: Device temperature
range from 10°C to 40°C permis-
sible!
Problem with internal voltage
SD card faulty!
<Info> button: Measurement with-
out data storage possible!
SD card faulty!
SD card inserted! Restart mea-
surement!
<Info> button: Storage of mea-
sured data possible after restart of
measurement!
SD card missing!
<Info> button: Insert SD card!
SD card missing!
<Info> button: Measurement with-
out data storage possible - or -
Insert SD card with correct ID!
SD card not readable!
<Info> button: Measurement with-
out data storage possible!
46
Cause
SD card with different ID to the
implant has been used
After the start of a continuous
measurement, data from another
implant have been received
Temperature in the Reader Unit
exceeds calibrated interval
Internal voltage of device too
high/too low
SD card not writeable or readable
(dirt, corrosion, faulty contacts)
SD card not readable (dirt, corro-
sion, faulty contacts)
Measurement starts without SD
card. During measurement, the
designated SD card for use with
that implant is inserted
No SD card inserted in data man-
agement mode
SD card not inserted
Insertion of wrongly formatted or
unformatted SD card
Reader Unit Set
Error detection / troubleshooting
Insert the matching M.scio SD card
into the Reader Unit. The implant
(see patient passport) is associ-
ated with the respective SD card
via the identification number (ID).
Increase the distance between the
two implants.
The Reader Unit can only be used
at device temperatures from 10°C
to 40°C. Current measurement is
stopped.
Reader Unit switches off automat-
ically after 20 seconds. Contact
Technical Service.
Use of implant's internal calibration
data. Data are not saved.
Check SD card for damage or dirt.
Start measurement again.
Insert SD card.
Insert the matching M.scio SD card
into the Reader Unit. The implant
(see patient data card) is associ-
ated with the respective SD card
via the identification number (ID).
Insert the matching M.scio SD card
into the Reader Unit. The implant
(see patient data card) is associ-
ated with the respective SD card
via the identification number (ID).