7.5. Battery Button

The battery button
allows you to check the status of the battery. Repeatedly pushing the
button will cycle through all the information.
• First, the gauge information will be shown
or
.
• Next
the number of charge cycles will be shown
or
.
After the third push of the battery button, the display will alternate back to the main screen
EN
showing the current flow setting. It will also automatically exit the battery status menu and
go back to the main flow setting display after approximately 15 seconds of no buttons being
pushed.

7.6. Responding to Alarms

CAUTION: If you are unable to hear or see alarms, do not have normal tactile sensitivity, or
cannot communicate discomfort, consult your clinician before using this device.
Pressing the alarm mute button
at any time will silence the buzzer. The length of the
mute period depends on the severity of the alarm (see Chapter 9 - Alarm Indicators). During
this mute period, the mute LED will remain illuminated, indicating the alarm buzzer is muted.
Push the mute button again to un-mute alarms. Pressing the mute button when there is no
active alarm will mute any future medium or low priority alarms for eight hours. See Chapter
9 - Alarm Indicators and Chapter 10 - Troubleshooting for additional information on alarms.
IMPORTANT: The alarm system is tested during the startup sequence. You should see all alarm
lights briefly turn on and the audible alarm indicator chirp. If alarms are suspected of mis-operat-
ing, contact your distributor for verification that alarms are working correctly.

7.7. Turning Off

CAUTION: Always turn off this device when not in use.
To turn the Zen-O lite™ Portable Oxygen Concentrator off, press and hold the power button.
The device will chirp and the screen will display a shutdown message
for ap-
proximately five seconds, then go into low-power mode.
IMPORTANT: Do not disconnect the AC power supply and remove the battery at the same time
while the unit is running. Always use the power button to turn the device off. Wait until the device
has completely shut down before disconnecting from power and removing the battery.