The date is displayed as the order of Year-Month-Day and Time of Hour-Minute (24-hour).

4.2 Alarm Setting

Note: Every time enter into the Alarm Setting in the main menu, you should input the
password.
Before setting, please enter password (1234) to set the parameter. Or you can direct access
to check the parameter but not to change.
How do you input passwords?
1. Press the Navigation button to change the numbers.
2. Press the OK button to confirm the number.
3. Press the Navigation button switch to the next numbers.
4. Press the OK button to confirm your selection.
5. Repeat the step one.
How to change the passwords?
In the measuring interface, press and hold the menu button for 5s to change the passwords.
The first time, enter the old passwords.
The second time, enter the new passwords.
Then you can change the passwords.
Notes:
1. The read-only password is 0000. Under this password, you can only check the parameter
but not change.
2. The make-changes password is 1234, enter this password, you can set the parameters.
3. You can change the password. If you forget the password, Please choose "Factory Default"
in System Setting, the password will Recover to factory password (1234) .
4. Every time enter into the Alarm Setting in the main menu, Alarm Volume and Alarm Pause
in the System Setting, you should input the password.
From the main menu, select and enter the "Alarm setting" screen, refer to Fig.4-4.
PULSE OXIMETER
Fig.4-3
- 18 -
INSTRUCTION MANUAL
You can select the Alarm on or off.
High limit SpO
range is71~100
2
Low Limit SpO
range is 70~99
2
High Limit PR range is 31~250
Low Limit PR range is 30~249
4.3 Data Management
From the main menu screen, select and enter the "Data Manage" screen, refer to Fig.4-5.
Press the Navigation button to select the sub-item to set, and then press the OK button to
confirm or Back button to return to the previous screen.
4.3.1 Data Review
Pick and enter the "Data review" interface as shown in Fig.4-6. By pressing the Navigation
button to review the records page by page.
The pulse oximeter can record the alarming parameter marked with red color. Press the Back
button, the pulse Oximeter returns to the previous interface.
MD300M
Fig.4-4
Fig.4-5
- 19 -