(4) Altitude Compensation (Alt
The fourth option of the "MENU" button is the Altitude Compensation option, used to adjust
pressure in the mask based on the altitude relative to sea level. Press the "UP" or "DOWN"
buttons to select from eight options (1 to 8) that correspond to a range of altitudes listed in the
table below. The level should be set relative to your elevation above sea level. Once the
preferred option has been selected, press the "MENU" button to confirm.
NOTE: Users can operate XT Fit at wide range of altitude within 0~8000 ft (limit to 5000 ft if
(5) Total Meter (TM
The fifth "MENU" item is the Total Meter which displays the total number of hours that the
device has been "On" and delivering airflow.
"START/STANDBY"
"START/STANDBY" button was pressed again to stop airflow. This meter should only be
re-set by the equipment provider, a respiratory therapist or by a physician.
(6) Compliance Meter (CM
Press "MENU" button to select [CM XXXX.X hr] menu, the compliance meter records the
total therapy hours for the device. The compliance meter should be re-set only by the provider,
a respiratory therapist or by a physician.
(7) Session Counter (CM
Press "MENU" button to select [CM XXX 4hr] menu, the session counter displays the
number of sessions greater than four hours. Your home care provider may ask for this
information. The session counter will be erased once the compliance meter has been reset.
the pressure is set at 18.5-20 cmH
accuracy pressure output at high altitude. Users must follow the below table to select
correct altitude setting depending on your elevation above sea level, the device will
automatically regulate airflow output to achieve targeted pressure at high altitude.
Otherwise, wrong altitude setting will cause inaccuracy pressure output.
There are eight altitude compensation levels. The level should be set depending on
your elevation above sea level. Below table is the comparison of altitude setting and
altitude.
Altitude Setting
1
2
3
4
5
6
7
8
XXXX.X)
button
XXXX.X)
XXX 4hr)
XXX)
O). The altitude function provides a method of
2
Altitude (Imperial)
0 ~ 1000 ft
1001 ~ 2000 ft
2001 ~ 3000 ft
3001 ~ 4000 ft
4001 ~ 5000 ft
5001 ~ 6000 ft
6001 ~ 7000 ft
7001 ~ 8000 ft
In other words, it is the time the
was
pressed
to
begin
9
Altitude (Metric)
0 ~ 304 m
305 ~ 609 m
610 m ~914m
915 m ~1219m
1220 m ~1524m
1525 m ~1829m
1830 m ~2134m
2135 m ~2438m
airflow