Starting the equipment

Place the scales on a secure, level base. Set the adjustable legs using the
spirit level to ensure the scales are mounted horizontally. Insert the power
adapter cable into the rear panel of the platform to connect power to the
personal scales.
Weighing
Step on the scales (you do not need to turn on the device).
Height measurement with BMI display
Carefully fold up the height rod and push it up. Place the top part on the
patient's head.
The display shows the size in centimetres followed by the body mass index.
Taring
The function key on the rear of the display resets the scales to zero.
Press the key. The display shows »0.0 kg« and the scales are ready for use.
Mother/child function
An adult stands on the scales.
> The display may show »47.6 kg« for example
Press the function key. > The display changes to »0.0 kg«
The mother can now step off the scales... > Display »–47.6 kg«
...and steps onto the scales again with the child.
> Now the display indicates the child's weight, for example »5.1 kg«
When the mother and child step off the scales, the display will show for
example »–47.6 kg«. Pressing the function key allows you to reset the
scales to »0.0 kg«.
The scales do not automatically reset to »0.0 kg« so that several tarings
can be performed in turn.
Switching the scales off
The scales automatically switch off when the display has shown »0.0 kg«
for approx. 15 seconds. To switch the scales off completely, remove the
mains adapter from the scales or the connector from the socket.
Calibration meter
The calibration meter reading shows how often the
scales have been calibrated. The meter reading stored
must match the calibration meter plate secured with an adhesive label.
Calibration meter reading display
The calibration meter reading is displayed for 2 seconds when the mains
power adapter is inserted.
Interface: USB
Can only be used for EN 60601-1 approved medical devices or for special
purposes
Getting started/Weighing + taring
Error messages
Zero setting
– Not idle > check ground stability
> Contact service department
Taring not possible
– Not idle > Stand still
– Negative weight reading
> Disconnect power briefly
Overload (more than 250 kg on the platform)
> Remove weight from the platform.
Low load reading
> Contact service department
The scales are set to zero when powered up
> Step off the scales > Start again > Contact service
department
Height rod not in reference position on power-up
> On power-up, fold in and insert the height rod fully.
Application part description
Service port
Option: USB port
Display
Stand
Spirit level
4 adjusting feet
below platform
ENGLISH
Reset key
Function key
(taring)
Mains adapter
(green LED lit,
mains supply present)
Weighing platform
GB
5